TROUBLESHOOTING






FLOWCELL LEAKS

A small amount of salts accumulating around the juncture between the auxiliary and working electrodes is normal. But any visible liquid or dripping should be dealt with. The most common reasons for this are:




HIGH PRESSURE

Pressures as high as 4000 psi are not necessarily a problem. However, a rise in pressure above what is typical for a given column and system may indicate a problem such as clogged tubing or an occluded column. If a high-pressure problem develops, look at the following:




LOW PRESSURE

As with high pressure, low pressure is only a problem if it is unusual for a given system and column. A drop in pressure can indicate a problem with the pump, or a massive leak in the system. Examine the following items if low pressure develops:




PRESSURE FLUCTUATION

Pressure fluctuations are a consequence of the alternating pulsations of the piston pump. They are minimized by good system design, such as the size and stroke length of pistons and the use of a pulse damper. Leaks and air bubbles are the major causes of excessive pressure fluctuation. Check the following items:




HIGH BACKGROUND

The background reading represents the electrochemical activity of the mobile phase, including all its components and contaminants. It is measured as the detector output when all offsets (zeroing) have been turned off. Every combination of mobile phase, electrode material, and applied potential has a characteristic background. The typical background current for your application is of diagnostic value — it should be noted each time the system is used and when mobile phase or hardware is changed. Backgrounds increase with temperature, applied potential, and contamination, and can decrease if the working electrode becomes coated or the reference electrode becomes depleted.

Many chromatographers don't understand the importance of a low background. They think that rezeroing gives them a zero background. But zeroing simply moves the frame of focus — the high background is still there, and it contributes to noise and distortion. This is true for both electrochemical and optical detectors.

Whether it be UV or EC detection, a high background makes it difficult to distinguish small peaks, as they represent only a tiny fraction of the total signal. When the background is low, however, these same small peaks are easier to distinguish because they represent a greater proportion of the total signal. High backgrounds can cause increased baseline noise, and are an indication that something is wrong with the system. Check the following if high backgrounds become a problem:




LOW BACKGROUND

Low backgrounds may seem desirable, but atypically low ones may indicate a problem. Be suspicious of a low background, especially when coupled with an unnaturally quiet baseline and small peaks. Examine the following items if backgrounds are low:




REGULAR BASELINE NOISE

Regular baseline noise (fluctuation) has a constant period that usually can be traced to a system component, especially if the period matches that of the pump. Air bubbles (and leaks) will produce baseline noise as they compress (or leak) at each pump stroke. To test whether the noise is flow related, change the pump speed. The period of the baseline noise should change proportionately. Consider the following items when tracking down baseline noise:




IRREGULAR BASELINE NOISE

Irregular baseline noise (fluctuation) can be difficult to track down. Sources can be internal or external to the LC system. Consider the following items:




BASELINE SPIKES

Spikes are fast deflections in the baseline, either positive, negative, or both. They usually occur at irregular intervals, but depending on their source can be regular as well. Evaluate the following possible causes:




BASELINE DRIFT

Baseline drift is a change in background over a long period of time, usually hours. A decrease in background after initially turning on the cell is normal, as the detector equilibrates. For other causes, consider the following items:




SMALL PEAKS

If peaks have normal shape and retention times but are smaller than expected, check the following:




NO PEAKS

Lack of peaks can be either a separation or a detection problem. Consider the following:




CONTAMINANT PEAKS

Contaminants are not necessarily a problem, unless they interfere with an analyte peak. If a contaminant does cause a problem, its source must be known before it can be dealt with. Common sources include the sample, chemicals used in extraction or purification, and mobile-phase components. Consider the following when tracking down a contaminant problem:




LATE ELUTERS

These are normal sample components that elute much later (even hours later) than the analytes of interest. Consider the following:




SPLIT PEAKS

A split peak is one peak that appears to be two poorly separated peaks. Consider the following:




BROAD PEAKS

The lowest detection limit is achieved when an analyte is in as compact a band as possible within the flow stream. We see this compactness as a narrow peak on the chromatogram. In isocratic elution peaks broaden as retention time increases, so it's best to elute the analytes as early as possible, consistent with good separation. Unusually broad peaks suggest the following:




RETENTION TIME CHANGES

Retention times can vary randomly, or steadily increase or decrease. Consider the following: